Catalog Number 121731048 |
Device Problems
Device Dislodged or Dislocated (2923); Noise, Audible (3273)
|
Patient Problems
Pain (1994); Synovitis (2094); Discomfort (2330); Joint Laxity (4526); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 03/24/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Revision procedure for patient experiencing pain.Xray revealed worn polyethylene liner.Operative notes indicated discomfort, loud squeaking, polyethyline fracture, fractured worn polyethyline with subluxation out of the ring fixation, wear debris and synovitis.Provided x-ray image and explanted device photographs confirms a rim fracture of the poly liner.Review of the x-ray and the femoral head photo finds sufficient evidence of a disassociation event.Doi: unknown; dor: (b)(6) 2021; left hip.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d4 (lot #, udi), d6a, d9, d10, g4 and h4.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d1, d2a, d2b and d4 (catalog).
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Visual examination of the provided x-ray image found evidence of disassociation.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
|