The ventilator was investigated on site and the o2 gas module was replaced and returned for further investigation.Simulated use test of the received o2 gas module has not reproduced the described customer issue with flow transducer failed during pre-use check.The review of the received device logs confirms the reported issue.The measured failed value in flow transducer test during pre-use check is just outside the accepted range.The failed pre-use check has been found mars 12, therefore the date of event was determined as (b)(6) 2021.If this fault happens during ventilation, the patient may receive an amount of oxygen in the gas mixture that deviate from the expected.Flow and pressure may also deviate from the expected.Alarms will be activated if the failure occurs during ventilation.The reported problem could not be reproduced, therefore the cause has not been established in this investigation.
|