Medtronic received information that during use, the customer reported air bubbles in the custom pack's fusion oxygenator during ecmo.
Use of the device was discontinued due to the patient's condition.
The patient subsequently died.
The reported device issue was thought by the customer to be associated with the patient's death.
Additional information: the customer thought the yellow air bubbles were plasma serum, that event was recorded through the gas out at the period time 14hr, 30hr, 36hr respectively since the system was working.
The recirculation port was used during priming/during the case.
Large amounts of suction and venting were not used.
There was no visible air in the system/tubing.
A bubble detector was not used.
The purge line was run open, attached to the port in the cardiotomy inlets.
|