MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE SCREWS; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE SCREWS; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number CTL4.5-040-S
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
The device has not been returned for evaluation as it remains in-situ.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event Description
Information was received that a revision procedure was performed in (b)(6) 2021 to address loose screws.As per the reporter, approximately two and a half weeks post-implantation, it was discovered that the screws in the proximal bone segment had loosened and disengaged the plate.No patient adverse event was reported.During the revision, the screws were re-tightened and the screw holes were back-filled with cement to prevent future screw back-out.Post revision, the patient has successfully resumed lengthening.Report 3 of 3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE SCREWS
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr, suite 100
aliso viejo, CA 92656
MDR Report Key11712092
MDR Text Key247012713
Report Number3006179046-2021-00254
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00887517029874
UDI-Public887517029874
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193617
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCTL4.5-040-S
Device Lot NumberACAEAD0826
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/26/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-