BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U128 |
Device Problems
Over-Sensing (1438); Pacing Problem (1439)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/14/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that this recently implanted cardiac resynchronization therapy pacemaker (crt-p) system, exhibited intermittent crosstalk oversensing, leading to ventricular pacing inhibition greater than two seconds.It was noted the patient was asymptomatic.Technical services (ts) reviewed device data and observed intrinsic r waves and recommended troubleshooting and reprogramming options.This patient was seen in clinic later the same day.The physician eliminated the crosstalk oversensing by reprogramming ventricular sensitivity.Patient monitoring to continue.No adverse patient effects were reported.The device remains in service.
|
|