Model Number 1054260
Device Problem Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/16/2021
Event Type  malfunction  
Event Description
Ventilator was returned to the manufacturer for service.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center, a "service required" code was found in the ventilator's downloaded error log.The device's internal battery was replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
adam price
312 alvin drive
new kensington, PA 15668
MDR Report Key11718434
MDR Text Key247081862
Report Number2518422-2021-01188
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959026353
UDI-Public00606959026353
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1054260
Device Catalogue Number1054260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2021
Date Manufacturer Received04/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-