Model Number 21-7302-24
Device Problems Excess Flow or Over-Infusion (1311); Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/21/2021
Event Type  malfunction  
Event Description
Information was received indicating that during bench testing of this smiths medical cadd cassette reservoirs, overdelivering at inconsistent rates was noticed.It was reported that when testing solis pump using the cadd cassettes with 20 ml of pca dose and checking the delivery percentage the cassettes, were overdelivering 2ml plus over our limit.No patient injury reported.
 
Manufacturer Narrative
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problems or issues were identified during this device history record review.The returned samples were received inside a plastic bag, in decontaminated condition without its original packaging.Visual and functional testing were performed.The sample was visually inspected at 12 to 16 inches under normal conditions of illumination according to procedure.Visual inspection did not present any damage, scuffs, pinch marks, cracks, crazing, etc.That could cause the failure mode reported.During the functional testing of the twenty-four (24) samples received only one (1) was tested to confirm the failure mode reported.The sample was fully primed and connected without difficulty, the pump was set running and the alarm was not activated, however delivery accuracy test results were out of specifications.The complaint was confirmed.The root cause cannot be determined.A corrective and preventative action has been opened to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD CASSETTE RESERVOIRS - FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key11721623
MDR Text Key247209042
Report Number3012307300-2021-03434
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586027239
UDI-Public10610586027239
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7302-24
Device Catalogue Number21-7302-24
Device Lot Number4084909
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-