ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number ASKU
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Inflammation (1932)
Event Date 03/20/2021
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported an "outbreak" of toxic anterior segment syndrome (tass) at their facility.In a follow up with the customer, it was reported that there were four patients who presented with tass following ocular surgery.There were two surgeon's involved.All four of the tass cases also had glaucoma and three of the patient's had combined cataract and glaucoma procedure.It is unknown at this time which surgery was performed on which patient.This file represents one of the four patients.
 
Manufacturer Narrative
The customer did not retain the product lot information for this procedure pack, therefore the device history records traceable to the reported procedure pack could not be reviewed.The customer reported that the their facility experienced an outbreak of tass and reported the cassette used as the source was unknown.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.In-process controls are established to ensure each final assembled cassette is verified that all required tests have been performed and all acceptance criteria are met prior to release.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key11728341
MDR Text Key247712929
Report Number1644019-2021-00283
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 07/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received06/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM; DUOVISC VISCOELASTIC SYSTEM; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Patient Outcome(s) Other;
-
-