MAQUET CRITICAL CARE AB SERVO-I; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number SERVO-I
Device Problem Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the ventilator failed pressure transducer test during pre-use check.There was no patient involvement.Manufacturer ref.#: (b)(4).
 
Event Description
Manufacturer's ref #: (b)(4).
 
Manufacturer Narrative
The ventilator was investigated by our field service engineer.The o2 gas module was replaced and the ventilator passed all functional and safety tests and was cleared for clinical use.The replaced part was not returned for investigation.No ventilator logs were provided.Since no parts were returned for investigation, the root cause of the event has not been determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVO-I
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
sara soderholm roman
roentgenvagen 2
solna 
MDR Report Key11730538
MDR Text Key247467737
Report Number8010042-2021-00929
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K123149
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSERVO-I
Device Catalogue Number6487800
Was Device Available for Evaluation? No
Date Manufacturer Received02/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2009
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-