MEDTRONIC NAVIGATION, INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Model Number 9736242 |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/26/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that the system would not boot up.Displayed, "rescue mode" after the medtronic logo, then would not proceed past "rescue mode" boot screen.The representative hard rebooted three times.Swapped outlets.Ctrl + alt + del done would just restart and boot back into the "rescue mode" there was no patient present.
|
|
Manufacturer Narrative
|
H3) a hardware analysis of 9736242 was initiated to determine the probable cause of the issue.Analysis found computer reported "error: failure.Computer beeping a beep code of 2 short beeps and one long beep repeatedly.Unable to enter in to operating system and key board was unresponsive.Isolated the problem to a defective ssd.Installed a known good ssd and the computer functioned normally without failure.The following codes applied to 9736242: fdm b01 fdr c02 fdc d02 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the (rep) and it was stated the new computer fixed the problem.
|
|
Manufacturer Narrative
|
H3) a medtronic representative went to the site to test the system.Hardware parts were replaced.The system then passed the system checkout and was found to be fully functional.The system performed as intended.Note: analysis of the replaced part was documented in the first follow up mdr.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|