Model Number BELLAVISTA 1000 US |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/02/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer reported that the bellavista 1000 us screen went black while on patient use.The customer confirmed that there was no patient harm associated with the reported event.
|
|
Manufacturer Narrative
|
Result of investigation: it was determined that the root cause of the issue was software communication error with internal o2 sensor (since g6).The unit start to work normally after the next restart at (b)(6) 2021 13:13:50.Installed software version is v6.0.1600.0.Prior to cfb logging "internal o2 sensor thread sampler" error log message is visible at (b)(6) 2021 13:07:28.Software bug in improved internal o2 sensor communication handling found in v6.0.1600.0.This issue is resolved with v6.0.1700.0.
|
|