STERIS MEXICO, S. DE R.L. DE C.V. CENTURY STERILIZER Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Superficial (First Degree) Burn (2685)
Event Date 03/17/2021
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite to inspect the century sterilizer and found that the pipe going from the generator to the steam manifold was contacting the side panel.This allowed the side panel to heat up and caused the reported event to occur.The technician made the necessary repairs, tested the unit, confirmed it was operating according to specification, and returned it to service.A 3-year complaint review indicates this to be an isolated event.No additional issues have been reported.
 
Event Description
The user facility reported that an employee obtained a burn on her upper arm after contacting the side panel of their century sterilizer.Medical treatment was sought and administered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURY STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key11733801
MDR Text Key260132212
Report Number3005899764-2021-00017
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/29/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-