MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 5MM/6MM SHORT TROCAR; PIN, FIXATION, THREADED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 5MM/6MM SHORT TROCAR; PIN, FIXATION, THREADED Back to Search Results
Catalog Number 71067307
Device Problems Break (1069); Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2021
Event Type  malfunction  
Event Description
It was reported that during an ankle external fixation procedure a 5mm/6mm short trocar (case: (b)(4)) and two 5mm/6mm tissue protector for sh half pin (case: (b)(4) and case: (b)(4)) were causing the drill to rip apart as they weren't wide enough for the 3.5 drill.The procedure was completed using back-up devices from smith and nephew; no delay was reported.No patient injury or other complications were reported.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device shows signs of extensive use.A functional evaluation was performed on the device and could not confirm the stated failure mode.The returned parts function as intended.A review of complaint history did not reveal additional complaints for the listed item.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5MM/6MM SHORT TROCAR
Type of Device
PIN, FIXATION, THREADED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11735802
MDR Text Key247684452
Report Number1020279-2021-03557
Device Sequence Number1
Product Code JDW
Combination Product (y/n)N
PMA/PMN Number
K120871
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71067307
Device Lot Number12BM19101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2021
Date Manufacturer Received05/19/2021
Is This a Reprocessed and Reused Single-Use Device? No