• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. ULTRAFLEX IAB: 7.5FR 40CC SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. ULTRAFLEX IAB: 7.5FR 40CC SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000260
Device Problem Fluid Leak (1250)
Patient Problem Needle Stick/Puncture (2462)
Event Date 07/14/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the patient had acute myocardial infarction and occlusion of the anterior descending artery. Acute pci operation was performed to open the criminal pathway. After the operation, an intra-aortic balloon (iab) was inserted for auxiliary protection. After the intra-aortic balloon pump (iabp) was started, blood was found in the helium tube. It was judged that the balloon had ruptured. As a result, the iab was pulled out, and another iab was inserted into the patient at a different insertion site. There was no report of patient complications or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameULTRAFLEX IAB: 7.5FR 40CC
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
2400 bernville road
reading, PA 19605
MDR Report Key11737208
MDR Text Key247689107
Report Number3010532612-2021-00111
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K000729
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial,Followup
Report Date 04/07/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/28/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date07/31/2020
Device Model NumberIPN000260
Device Catalogue NumberIAB-06840-U
Device Lot Number18F18H0050
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/23/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-