• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. OSS AVL TIB BEARING 12MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. OSS AVL TIB BEARING 12MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Failure of Implant (1924); Pain (1994)
Event Date 08/14/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0001825034-2021-01278, 0001825034-2021-01279, 0001825034-2021-01280, 0001825034-2021-01282, 0001825034-2021-01283, 0001825034-2021-01284.Concomitant medical products: oss avl tibial lock ring, item# 161073, lot# 466660.Oss avl yoke 12mm, item# 161075, lot# 382130.Oss avl mod tib plate oss tape, item# 161065, lot# 690080.Oss tib blk aug 10x63/67 univ, item# 150426, lot# 105590.Oss avl poly tib bushing set, item# 161071, lot# 490920.Oss poly femoral bushings, item# 150477, lot# 499870.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient was revised approximately five months post-implantation due to locking nut fractured in two pieces and allowed piston to disengage from the tibial component.Attempts to obtain additional information have been made; however, no more information is available.
 
Manufacturer Narrative
Upon further investigation, it was determined that this item was reported in error.The locking ring fracture is what caused the yoke to disassociate from the tibial tray.Therefore, the locking ring is the only reportable item.The initial report was submitted in error and should be voided.
 
Event Description
Upon further investigation, it was determined that this item was reported in error.The locking ring fracture is what caused the yoke to disassociate from the tibial tray.Therefore, the locking ring is the only reportable item.The initial report was submitted in error and should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSS AVL TIB BEARING 12MM
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11739339
MDR Text Key247772745
Report Number0001825034-2021-01281
Device Sequence Number1
Product Code KRO
UDI-Device Identifier00880304456136
UDI-Public(01)00880304456136(17)230907(10)809890
Combination Product (y/n)N
PMA/PMN Number
K051570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,health p
Type of Report Initial,Followup
Report Date 05/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/07/2023
Device Model NumberN/A
Device Catalogue Number161068
Device Lot Number809890
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-