Model Number N/A |
Device Problems
Flaked (1246); Fracture (1260); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/12/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has been returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the flexible g7 drill driver silicone wrap broke and is peeling off.In the efficiency g7 pan.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d4; d9; g3; h2; h3; h4; h6. one flexible silicone drill driver was returned and evaluated.Upon visual inspection the device sheath had torn exposing the inner spring.A review of the device history records identified no deviations or anomalies during manufacturing.The device is used for treatment.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|