Information was received regarding a cadd administration set.It was reported that the writer presented to the client's house to change a cadd pump tubing.The tubing cassette was loaded into the cadd pump and the med bag spiked.Priming was initiated and the pump was sounding like it was drawing fluid, but nothing was coming into the tubing.The cassette was removed and reloaded, and user ensured the spike was fully into the med bag.The batteries were changed to ensure enough power to pull the medication through, however, there was still no fluid being pulled through the tubing.The write notified the gnh hpt program, but no solutions were provided.The writer provided the patient with new tubing from a new lot and then primed the new bag without any incidents.There was no adverse effects reported.
|