• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN PURITAN BENNETT; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN PURITAN BENNETT; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 980
Device Problem Device Displays Incorrect Message (2591)
Patient Problem Insufficient Information (4580)
Event Date 02/03/2021
Event Type  malfunction  
Event Description
Ventilator screen started to say "patient disconnect error" while the patient was connected to the ventilator.Rn and rt attempted all troubleshooting options; however, ventilator continued to report "patient disconnect error".Patient removed from ventilator and placed on a different vent.Biomed made aware.Ventilator taken out of service and taken to biomed department.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURITAN BENNETT
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
MDR Report Key11744378
MDR Text Key247921523
Report Number11744378
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/21/2021,04/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number980
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/21/2021
Date Report to Manufacturer04/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-