Model Number SERVO-S
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2021
Event Type  malfunction  
Manufacturer Narrative
During investigation on-site performed by our field service engineer presence of moisture/corrosion/liquid spill in the printed circuit board (pcb) was found.The pcb were replaced and after successful tests the ventilator was sent back in service.The pcb were not further investigated since ingress of liquid/corrosive substance on printed circuit boards can cause failure/short circuits which might lead to a ventilator malfunction.There is no information on how the liquid spill entered the ventilator or what kind of liquid spill it was.
 
Event Description
It was reported that the ventilator failed flow transducer test during pre-use check.There was no patient involvement.Manufacturer ref.#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVO-S
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
sara soderholm roman
roentgenvagen 2
solna 
MDR Report Key11744493
MDR Text Key248784317
Report Number8010042-2021-00967
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K123149
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSERVO-S
Device Catalogue Number6640440
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2009
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-