• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC TRILOGY EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number DS2110X11B
Device Problems Circuit Failure (1089); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2021
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator inoperative condition occurred.The device was not in patient use.The ventilator was returned to the manufacturer for evaluation and the customer's complaint was confirmed.The device's system board was replaced to address the issue.During the evaluation of the device at the manufacturer's service center, an issue related to the device's display was observed.The device's display was replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC
1001 murry ridge rd
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC
1001 murry ridge rd
murrysville PA 15668
Manufacturer Contact
adam price
312 alvin dr.
new kensington, PA 15068
7243349303
MDR Report Key11744833
MDR Text Key247939258
Report Number2518422-2021-01248
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959051942
UDI-Public00606959051942
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181166
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDS2110X11B
Device Catalogue NumberDS2110X11B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2021
Date Manufacturer Received04/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-