ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE; UNIT, PHACOFRAGMENTATION
|
Back to Search Results |
|
Catalog Number 8065752914 |
Device Problem
No Device Output (1435)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that during a surgical procedure the performance of the handpiece is lost.The handpiece passed prime tune.Procedure details and patient impact were not reported.Additional information has been requested but not received.
|
|
Manufacturer Narrative
|
Additional information has been provided in d.9., h.3., h.6.And h.10.No further information was provided from this customer.The handpiece was returned for evaluation.The handpiece was received for evaluation.A visual assessment of the returned sample found no visual non-conformities.The handpiece was connected to a resistance test box, where it¿s input and output impedance were found to be within specification.The returned sample was connected to a calibrated system.The handpiece tuned successfully and completed a five-minute burn-in test with the system set at 100% ultrasonic and torsional power.This handpiece was investigated and subsequently dispositioned to scrap.The returned handpiece was found to exhibit no problems.Therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
|
|