• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX M100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX M100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 106697
Device Problems Disconnection (1171); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone number: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that part of a prismaflex m100 set had spontaneously detached at the joint above the filter before the deaeration chamber spraying blood in the patient room during treatment.The amount of blood loss was not reported.Air was observed in the filter and the lines were immediately clamped and was stopped.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; however, a picture of the impacted set was provided.The visual inspection did not identify any specific issue on the set.Nevertheless, the reported condition is considered as confirmed.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M100
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11752242
MDR Text Key249284855
Report Number8010182-2021-00152
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414064556
UDI-Public(01)07332414064556(10)
Combination Product (y/n)Y
PMA/PMN Number
K041005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Catalogue Number106697
Device Lot Number20B2323M
Was Device Available for Evaluation? No
Date Manufacturer Received06/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NI.
-
-