Model Number 2178-63-136 |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 04/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that while attempting to prepare the femur for a femoral sleeve, the adapter, shaft and subsequently t-handle were "slipping".Upon inspection after the procedure, wear is noted on each and will be sent for replacement.It was also reported that a separate form was submitted for the implants.With a 2 minutes delay in surgery.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : examination of the returned device confirmed the reported event.The investigation attributed the root cause of the event to device damage/wear from normal use and servicing and the need for corrective action was not indicated.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Search Alerts/Recalls
|