BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U128 |
Device Problem
Battery Problem (2885)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 01/31/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited sensing of chronic high atrial rates which may impact battery longevity over time.Please refer to the description for more information regarding the specific circumstances of this event.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) had six years remaining and from last device check it showed nine years.Moreover, the patient is im atrial fibrillation (af) recently.An engineering analysis was performed in which it showed that the device had high rate atrial sensing that caused an increase in power consumption.A device reprogramming was then recommended.To date, the device remains in service.No adverse patient effects were reported.
|
|