SYNTHES GMBH SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Model Number 532.110
Device Problems Noise, Audible (3273); Intermittent Loss of Power (4016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.The small battery drive device was evaluated and the reported condition that the device had internal rattling, and the speed of the device was inconsistent was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device had insufficient/low power, a worn motor, and worn bearings.It was further determined that the device failed pretest for check the power with the test bench.The assignable root cause of these conditions was determined to be traced to component failure due to normal wear.Udi ¿ (b)(4).
 
Event Description
It was reported that during pre-surgery testing for an open reduction internal fixation (orif) surgical procedure for the right elbow, it was observed that the small battery drive device had internal rattling while in use, and the speed of the device was inconsistent.It was reported that there was no delay in the procedure due to the event as an unspecified spare device was available for use.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE II
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kara ditty-bovard
eimattstrasse 3
oberdorf 4436
SZ   4436
6107428552
MDR Report Key11769160
MDR Text Key248792581
Report Number8030965-2021-03324
Device Sequence Number1
Product Code HWE
UDI-Device Identifier10886982241191
UDI-Public10886982241191
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 04/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number532.110
Device Catalogue Number532.110
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2021
Date Manufacturer Received04/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-