DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported issue.The screen was not displaying any advisory messages or alarming any power up test fault codes in normal operation mode.Power up test routine passed, system test was ok.Ac power and battery charging was functioning.The console was fully engaged inside the hospital cart.However, the event logs revealed user interface system failure of "fault code # 137 - video driver fault" logged 3 times.A new upper display monitor board & video generator board was installed to address the video driver fault errors.The fse then completed a preventive maintenance (pm)with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) shutdown.No patient harm, serious injury or adverse event was reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, b6, b7, g4, g7, g8, h2, h4, h6 (type of investigation), h10, h11 corrected field: g1 (contact person ¿ mfg site).
|
|