Model Number 0684-00-0568-01
Device Problem Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2021
Event Type  malfunction  
Manufacturer Narrative
Occupation: nurse manager event site postal code: (b)(6). Event site state - (b)(6). The device has not been returned to the manufacturer so we are unable to complete an evaluation. If provided we will send a supplemental report with our additional findings. Complaint record id # (b)(4).
 
Event Description
It was reported that during insertion of the intra-aortic balloon (iab) there was resistance encountered when the iab was inserted halfway into the sheath. The iab was removed from the sheath and it was discovered that the iab was broken. A new iab was inserted. There was no patient harm or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSENSATION PLUS 7.5FR. 40CC IAB
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
dorota wolpiuk
15 law drive
fairfield, NJ 
MDR Report Key11776220
MDR Text Key249024509
Report Number2248146-2021-00286
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K122628
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 08/16/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0684-00-0568-01
Device Catalogue Number0684-00-0567
Device Lot Number3000126872
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/07/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Initial

-
-