Model Number 2307-95-000 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/26/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the glenosphere orientation guide tip was bent during glenosphere implantation.There were no broken pieces and no delay in the procedure.Instrument will be returned to distributor a1a.There is no further information available.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned for evaluation.The investigation could not draw any conclusions about the reported event without the device to examine.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
|