DATEX-OHMEDA, INC. AESPIRE 7900 ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1009-9012-000
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/08/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Patient information is currently not available. A ge healthcare service representative performed a checkout of the system and confirmed the reported issue. The adjustable pressure limit (apl) valve was replaced to resolve the reported issue. Legal manufacturer: (b)(6).
 
Event Description
The hospital reported that the adjustable pressure limit (apl) valve would allow pressure to increase above setting, then would suddenly release pressure. There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAESPIRE 7900
Type of DeviceANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
joseph seliga
3000 n grandview blvd.
waukesha, WI 53188
MDR Report Key11792699
MDR Text Key249767176
Report Number2112667-2021-01206
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/07/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1009-9012-000
Device Lot NumberANCU00806
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received04/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/20/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-