SYNTHES GMBH SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.110 |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once the evaluation has been completed, a supplemental medwatch report will be sent accordingly. udi: (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that the small battery drive device had a grinding sound while in use.It was noted that the failure occurred more in forward mode.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown but was noted to have occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During evaluation it was determined that the device passed all manufacturing specifications.Therefore, the reported condition could not be duplicated nor confirmed.An assignable root cause was not determined.D9.The date the device was returned to the manufacturer was reported as may 6, 2021 in the initial report and has been updated to may 13, 2021.
|
|
Search Alerts/Recalls
|
|
|