• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVC REPL MDU HAND CNTRL PWRMX; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVC REPL MDU HAND CNTRL PWRMX; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616S
Device Problems Mechanical Jam (2983); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and observed no issues.A functional evaluation revealed a jammed motor/blade stall.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The complaint was confirmed and the root cause has been associated with a mechanical component failure.Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out.No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that during set up of a knee arthroscopy procedure, the mdu was plugged in but it made a bad grinding sound.No delay and a back up was available to complete the procedure.No other complications were reported.Results of investigation have concluded that this unit had a jammed motor/blade stall error which makes it a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC REPL MDU HAND CNTRL PWRMX
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11805145
MDR Text Key249971929
Report Number1643264-2021-01779
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010565105
UDI-Public03596010565105
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200616S
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/06/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-