SMITH & NEPHEW, INC. SURESHOT TARGETER; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 71692801
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2021
Event Type  Injury  
Event Description
It was reported that the host computer screen for the sureshot targeter displayed an "aiming sureshot is damaged, please replace with another one" message.It is unknown when the event happen relative to surgery.A changed in surgical technique was required to complete the procedure.It is unknown if there was a surgical delay.No patient injuries were reported.
 
Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device shows signs of extensive use.A function evaluation of the returned device confirmed the stated failure mode.After plugging the device into a monitor, the error message, "sureshot targeter - invalid or broken tool please exchange," appeared.A review of the complaint history for the listed part revealed prior complaints for the listed failure mode but no other complaints with the same serial number as this is a unique serial device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESHOT TARGETER
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11805773
MDR Text Key249964240
Report Number1020279-2021-03984
Device Sequence Number1
Product Code OLO
UDI-Device Identifier03596010648167
UDI-Public03596010648167
Combination Product (y/n)N
PMA/PMN Number
K092497
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71692801
Device Catalogue Number71692801
Device Lot NumberND3818
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2021
Date Manufacturer Received07/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other; Required Intervention;
-
-