• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955467
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2021
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment with a prismaflex m150 set c, an external blood leak was observed.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; however a picture was received and evaluated.The reported condition was not verified.The cause could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
B5: it was reported that the event was an internal blood leak and not an external blood leak.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Correction: b4/f8: date of this report in mdr follow up #2 is being corrected from blank to 06/21/2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M150
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11811853
MDR Text Key250826487
Report Number8010182-2021-00163
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414123451
UDI-Public(01)07332414123451
Combination Product (y/n)Y
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Catalogue Number955467
Device Lot Number20K2402
Was Device Available for Evaluation? No
Date Manufacturer Received07/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-