• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE TRACKING CAMERA; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE TRACKING CAMERA; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10025
Device Problem Intermittent Communication Failure (4038)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/25/2021
Event Type  malfunction  
Event Description
It was reported that the brand new real intelligence tracking camera was faulty out of the pelican case.Additionally, the camera was emitting a single beep at intervals in sync with a flashing orange light at the front of the camera and a green light (single dot) at the back of the camera was also flashing.No case involved; therefore, there was no patient involvement.
 
Manufacturer Narrative
H3, h6: the cori camera, part number rob10025, s/n (b)(6), intended for use in treatment was not returned for evaluation thus, a visual and functional evaluation could not be performed and a relationship between the reported event and the device could not be determined.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed a factor that may have contributed to the reported symptom may have been the enabled shock sensor.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE TRACKING CAMERA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11814265
MDR Text Key250524336
Report Number3010266064-2021-00366
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757437
UDI-Public00885556757437
Combination Product (y/n)N
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10025
Device Catalogue NumberROB10025
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-