Model Number TPL0059
Device Problem Imprecision (1307)
Patient Problem Discomfort (2330)
Event Date 04/15/2021
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment. Testing revealed that no failures were found and the system was working as intended. Analysis results for the software exports were not available as of the date of this report. A follow up report will be submitted when analysis is complete. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a guidance system being used during a spinal procedure. It was reported that two segments were done during a t10-s2 open fusion case. The first segment was t10-l3 and a dual clamp and bone mount bridge was used to mount the surgical system to the patient. The second segment was l4-s2 and a schanz arm and pin were used to mount the surgical system to the patient. The manufacturer representative noted that a drill was used when placing screws during the procedure. Imaging was done during the case and the surgeon thought the screws were placed well. Four days after the procedure, the surgeon told the representative that the s2 screw on the right side was deviated more than 10 mm. The patient experienced slight discomfort after the procedure. The representative indicated that the first 30 mm of the entry point looked identical to the planned trajectory, but the tap looked to have taken the path of least resistance down the joint. The surgery was delayed less than an hour.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMAZOR X STEALTH EDITION
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IS 3079567
Manufacturer (Section G)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IS 3079567
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key11819865
MDR Text Key250478186
Report Number3005075696-2021-00053
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 06/09/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/13/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberTPL0059
Device Catalogue NumberTPL0059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-