Model Number URF-P7 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device has not been returned to omsc but was returned to olympus france.(ofr).Ofr sent the device to a third party laboratory for microbiological testing.As a result of the testing, no microbe was detected from the sample collected from the all channels of the device.The testing result cleared the french guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that as a result of routine microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device.Bacillus cereus (2 cfu) and pseudomonas spp.(1 cfu) other detailed information such as the reprocessing method was not provided.There was no report of infection associated with this report.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.The subject device was inspected, and no defects or lack of conformity were found.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.If additional information becomes available, this report will be supplemented.
|
|