DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Failure to Calibrate (2440)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge service territory manager (stm) was dispatched to investigate and upon arrival, the stm looked through the logs and found no technical errors.The stm then performed a unit checkout.Unit passed all functional and safety tests.The stm informed that was not allowed to do a battery run time test.The iabp was returned to the customer and it is the customer's responsibility to run the battery test prior to use.
|
|
Event Description
|
It was reported that during check, the cs300 intra-aortic balloon pump (iabp) was unable to calibrate optical sensor.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Analysis of production: (3331) the device history record review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110) the overall 24 month product complaint trend data for the period may-2019 through apr-2021 was reviewed.There were no triggers identified for the review period.
|
|