Catalog Number ASKU
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Inflammation (1932); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer six patients experienced toxic anterior segment syndrome (tass) following a cataract extraction procedure's.A completed questionnaire was received indicating the patient experienced iris prolapse was during surgery.The patient required additional medical treatment with compounded antibiotic, steroid and non steroidal drops in the right eye.Current patient condition was not reported.This report represents patient five of the six patients.
 
Manufacturer Narrative
The customer did not retain the product lot information for this custom pak, therefore the device history records traceable to the reported procedure pack could not be reviewed.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted.A custom-pak is an assemblage of single-use medical devices and accessories provided to the customer in a sterile manner.An evaluation is conducted to ensure each pak and item in the pak meets customer, process, and regulatory requirements.Several sizes are available to accommodate the best fit for the custom-pak.Once the custom-pak is assembled and sealed, they are all sterilized as a complete unit.The sterilization validation has taken into account the worst case of the pouches to ensure all custom-paks meet all sterilization cycle parameters for acceptability prior to release.The root cause of the customer's complaint could not be established as a sample has not been received and therefore the condition of the product could not be verified.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.Custom pak and the components inside of the paks are single-use items provided to the customer in a sterile manner.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM-PAK SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key11827154
MDR Text Key250863671
Report Number1644019-2021-00339
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 08/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/14/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received08/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
BSS WITH EPINEPHINE 1:1000; DUOVISC; INFINITI VISION SYSTEM; MALYUGUN RING; MIOCHOL; PANOPTIX TORIC IOL; PHENYLEPHRINE 2.5; POLYMYXIN; PROPARACAINE; TROPICAMIDE 1%
Patient Outcome(s) Required Intervention;
-
-