Model Number 0998-00-0800-53 |
Device Problem
Failure to Calibrate (2440)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit.The fse replaced the video generator pcb and touchscreen assembly.The fse completed all safety, functionality and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.The first name of the initial reporter has been abbreviated due to field character limit; the full name should read (b)(6).The initial reporter is a getinge employee who has different contact details from that of the event site.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse) the touchscreen of the cardiosave intra-aortic balloon pump (iabp) would not calibrate.There was no patient involvement, and no adverse event reported.
|
|