Model Number ROSA ONE |
Device Problems
Connection Problem (2900); Unintended Application Program Shut Down (4032)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been evaluated yet for investigation purpose.Once the evaluation is performed, a follow-up medwatch report will be submitted.Unique identifier (udi) #: (b)(4).
|
|
Event Description
|
It was reported that a patient underwent an initial dbs procedure on (b)(6) 2021.Subsequently, there was a random shutdown of the (b)(4) and a failure to connect the controller.There was an approximate seven (7) minute delay.
|
|
Manufacturer Narrative
|
The analysis of the system logs indicates that during the event date an initialization error of the network card by rtx was detected.This issue is known and might correspond to the connection failure to the arm described in the complaint.However, information and data provided are insufficient to confirm the issue and its cause.The event described in the complaint cannot be confirmed.
|
|
Event Description
|
It was reported that a patient underwent an initial dbs procedure on (b)(6) 2021.Subsequently, there was a random shutdown of the rosanna software and a failure to connect the controller.There was an approximate 7 minute delay.
|
|