• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/23/2021
Event Type  malfunction  
Manufacturer Narrative
At this time, the customer has not requested getinge to evaluate the iabp.Additional information was requested from the customer with regard to the repair and status of the iabp; however, despite our best efforts, no repair information and no status of the iabp has been received.If any pertinent information is received in the future, a supplemental report will be submitted.Service not requested.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) began alarming "low helium" although the helium gauge showed almost full.The customer changed the helium tank with no change in the gauge or the alarm.The iabp was in rescue mode and the tank was opened to the iabp.The iabp was labeled and taken to biomed.No patient harm, serious injury or adverse event was reported.
 
Manufacturer Narrative
Updated fields: b4, b6, b7, d11, g4, g7, g8, h2, h6(investigation type), h10, h11 corrected fields: g1(contact person), h4.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key11841202
MDR Text Key251193438
Report Number2249723-2021-01038
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/09/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/07/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
Patient Weight119 KG
-
-