DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/23/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
At this time, the customer has not requested getinge to evaluate the iabp.Additional information was requested from the customer with regard to the repair and status of the iabp; however, despite our best efforts, no repair information and no status of the iabp has been received.If any pertinent information is received in the future, a supplemental report will be submitted.Service not requested.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) began alarming "low helium" although the helium gauge showed almost full.The customer changed the helium tank with no change in the gauge or the alarm.The iabp was in rescue mode and the tank was opened to the iabp.The iabp was labeled and taken to biomed.No patient harm, serious injury or adverse event was reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, b6, b7, d11, g4, g7, g8, h2, h6(investigation type), h10, h11 corrected fields: g1(contact person), h4.
|
|
Event Description
|
N/a.
|
|