• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US QUICKSET ACE GRATER HEAD 79MM; HIP INSTRUMENTS : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US QUICKSET ACE GRATER HEAD 79MM; HIP INSTRUMENTS : REAMERS Back to Search Results
Model Number 2440-00-579
Device Problem Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/10/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the graters size 36-80 are dull and need to be sharpened.No patient consequences or surgical delays reported.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned.The investigation could not verify or identify any product contribution to the reported event with the information provided.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre) was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET ACE GRATER HEAD 79MM
Type of Device
HIP INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11849574
MDR Text Key251444289
Report Number1818910-2021-10694
Device Sequence Number1
Product Code HTO
UDI-Device Identifier10603295124269
UDI-Public10603295124269
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2440-00-579
Device Catalogue Number244000579
Device Lot NumberSO2011668
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2021
Date Manufacturer Received06/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-