Model Number N/A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/03/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical products: item number 31-301322 item name arcos con sz b std 70mmtrl lot # 427427.Item number 11-300817 item name arcos 17x150mm spl tprdist lot # 467680.Item number 31-301850 item name arcos mod t handle lot # unk.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the when the surgeon went to tighten the cone body trial onto the distal sts 17x150 stem and the tip of the 3.5 torque limiting screw driver broke into the cone body trial.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
  one stem, taper body, and fractured driver were returned and evaluated.Upon visual inspection the body and stem were stuck together and could not be separated.The fractured piece of the driver that was returned is stuck inside of the body and could not be removed.Lot identification is necessary for review of device history records, lot identification was not provided.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOS 3.5MM HEX DRIVE
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11855472
MDR Text Key251695830
Report Number0001825034-2021-01560
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K090757
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number31-301852
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-