Model Number N/A |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/03/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Concomitant medical products: item number 31-301322 item name arcos con sz b std 70mmtrl lot # 427427.Item number 11-300817 item name arcos 17x150mm spl tprdist lot # 467680.Item number 31-301850 item name arcos mod t handle lot # unk.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the when the surgeon went to tighten the cone body trial onto the distal sts 17x150 stem and the tip of the 3.5 torque limiting screw driver broke into the cone body trial.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
one stem, taper body, and fractured driver were returned and evaluated.Upon visual inspection the body and stem were stuck together and could not be separated.The fractured piece of the driver that was returned is stuck inside of the body and could not be removed.Lot identification is necessary for review of device history records, lot identification was not provided.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|