MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 STATLOCK SLIDING POST CRESCENT PAD SKIN PREP; STAND, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 STATLOCK SLIDING POST CRESCENT PAD SKIN PREP; STAND, INFUSION Back to Search Results
Catalog Number PIC0220CE
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/14/2021
Event Type  malfunction  
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a detached retainer is confirmed and was determined to be manufacturing related.Two picc plus statlock kits were returned for evaluation.An initial visual observation showed one of the kits was returned open and one kit was returned sealed.The retainer of the statlock device returned in the open kit was found to be completely detached, and the underside of this detached retainer was observed to be tacky and stuck slightly to the edge of the detached pad.The detached pad was also found to be tacky at this location.The underside of the retainer as well as the pad of the sample returned in the sealed kit were also found to be tacky.A microscopic observation revealed good adhesive coverage on the underside of both retainers; however, strings of adhesive were found on the underside of the retainer of the sealed sample when an attempt was made to remove it from the pad.This suggests that the adhesive between the retainer and pad may not have been fully cured.The investigation has been forwarded to the manufacturing facility for further evaluation, and bd is working closely with the manufacturing facility to prevent recurrence of the reported event.A lot history review (lhr) of juer1322 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (juer1322) have been reported from multiple facilities in (b)(4).
 
Event Description
It was reported the plastic clamps on the statlock device falls off the band aid so the picc catheter is at risk for sliding out off the patient because it is not fixed.So plastic clamps are not fixed correct on the band aid.No other information was provided.On (b)(6) 2021- two devices were returned.This report addresses the second device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATLOCK SLIDING POST CRESCENT PAD SKIN PREP
Type of Device
STAND, INFUSION
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DAVOL SURGICAL INNOVATIONS -9616067
ave. roberto fierro #6408
parque industrial aeropuerto
cd. juarez, chih s.a. de c.v. 32690
MX   32690
Manufacturer Contact
kelsey erickson
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key11856200
MDR Text Key252605217
Report Number3006260740-2021-01899
Device Sequence Number1
Product Code FOX
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPIC0220CE
Device Lot NumberJUER1322
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-