Model Number 9735665 |
Device Problems
Mechanical Problem (1384); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 9735821, serial/lot #: (b)(4).The camera was returned for analysis.As reported, the left detector lens was missing.The returned position sensor unit (psu) also had many nicks and scratches on the housing.A check of the event log showed intermittent firmware incompatibility in (b)(6) 2020, intermittent low illuminator current in sep 2020, and intermittent sensor not functioning in (b)(6) 2021.Otherwise, the psu passed an accuracy test (aak) at.08 mm with a passing threshold of.25 mm.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the camera had the cap attached at the end of a surgery that finished at midnight.When the representative returned in the morning after moving the navigation system to a different part of the operating room, the cap was missing and cannot be found.No patient present.Additional information noted that they have seen the occasional lens cap that has fallen out, not uncommon.
|
|
Manufacturer Narrative
|
H2) correction: h3 corrected.H3: a medtronic representative went to the site to test the equipment.Testing revealed that the camera was required to be replaced.Once replaced, the imaging system then passed the system checkout and was found to be fully functional.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|