Model Number 72200616S
Device Problems Mechanical Jam (2983); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that the mdu had an e6 error code.No case reported; therefore, there was no patient involvement.
 
Manufacturer Narrative
H3, h6: the reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A relationship, if any, between the subject device and the reported event could not be determined.No containment or corrective actions are recommended at this time.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A complaint history review concluded this was a repeat issue.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.There was a relationship found between the returned device and the reported event.A visual inspection was performed on the exterior of product and no physical damage was observed.Product failed functional testing with blade stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The gearbox was found to be jammed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed, and the root cause was associated with corrosion in the gearbox.Factors which can contribute to overheating include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This can be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC REPL MDU HAND CNTRL PWRMX
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11858100
MDR Text Key252239577
Report Number1643264-2021-01920
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010565105
UDI-Public03596010565105
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200616S
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received06/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-