Model Number CLK-4
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to the service center for evaluation.The evaluation found the spare lamp does not ignite.Additionally, the reported front power switch cover is damaged was confirmed.Also, the front panel is damaged/cracked and the air supply is weakened due to a defective pump.The device is pending repair.A review of the repair history shows no previous repair records; the device was purchased on (b)(6) 2013.The investigation is ongoing; therefore, the root cause of the reported malfunction cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The technical assistance group was informed, during an unspecified event the halogen light source reportedly had a broken power switch cover.No patient injury or harm was reported.
 
Manufacturer Narrative
The supplemental report is submitted to provide the result of the legal manufacturer¿s investigation.The device history record for the subject device was reviewed and it was determined that no anomalies were found that could have caused or contribute to the reported problem.The legal manufacturer performed an investigation, a conclusive root cause was not identified.The device has been in operation for more than eight years.It was determined a potential cause could be related to due to long-term repeated use, or that the harness has corroded due to aging, resulting in poor lighting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11860908
MDR Text Key251945355
Report Number8010047-2021-06566
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170024528
UDI-Public04953170024528
Combination Product (y/n)N
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 06/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLK-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2021
Was the Report Sent to FDA? No
Date Manufacturer Received06/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-