Catalog Number UNKNOWN |
Device Problem
Difficult or Delayed Activation (2577)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Manufacturer ref# (b)(4).Blank fields on this form indicate the information is unknown or unavailable.Catalog# is unknown but referred to as unknown femoral filter.Investigation is still in progress.
|
|
Event Description
|
Description of event according to initial reporter: unknown femoral filters.Complaint is that they've had trouble deploying the filters.Release buttons get stuck.Devices are not letting go of all the legs.
|
|
Manufacturer Narrative
|
Manufacturer ref# (b)(4).Summary of investigational findings: it was reported that an unknown femoral filter device (suspect celect-pt) had trouble deploying the filter.Release button got stuck and the device didn¿t let go of all the filter legs.No patient harm indicated so it is assumed that the filter was placed correctly eventually.Cook was unable to conduct a review of the device history record, as the lot number of the complaint device was not provided for the investigation.According to the instruction for use excessive force should not be exerted in placement of the filter.Based on the limited provided information it is unknown what caused the reported event.It is possible that during the release of the filter excessive force was used which resulted in the described event, but this is only speculation.No evidence to suggest that the device is not manufactured according to specification cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
|
Search Alerts/Recalls
|