Model Number 71674067
Device Problems Material Deformation (2976); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/04/2021
Event Type  malfunction  
Event Description
It was reported that, during a trauma procedure, it was noticed that the following instruments of the trigen¿ intertan¿ instrument set are warped due to wear and tear over past 3 years: lag screw drill sleeve ((b)(4)), intertan 3.2mm guiide pin sleeve ((b)(4)), lag screw tap ((b)(4)), lag screw driver ((b)(4)), compression screw hexdriver ((b)(4)), subtroc lag screw driver ((b)(4)) 130 rad drill gde drop ((b)(4)), unknown trigen instrument ((b)(4)), unknown trigen instrument ((b)(4)).The procedure was able to be completed with the same device.Surgery was greater than 30mins delayed.Patient was not harmed.
 
Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned lag screw driver confirms the device is warped at the ends, rendering it inoperable.This device also has excessive wear usage.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREW DRIVER
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11861860
MDR Text Key252000053
Report Number1020279-2021-04563
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560605
UDI-Public03596010560605
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674067
Device Catalogue Number71674067
Device Lot Number07FSP0012
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2021
Date Manufacturer Received09/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-