DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate, the fse verified that the low level connector was not working.Once the fse looked at the connector internally, he found that the connector pins were broken away from the front end board.He tried replacing the board and found that with the new front end board, the iabp was alarming start up failure 12.He removed the board and ordered a new replacement board.The fse returned and replaced the board again and had the same issue with the new front end board.He removed the board and checked the pins because both boards were somewhat difficult to insert.The fse found that the pins on the boards were bent and one was broken.He then checked the connector on the back plane board and noticed that some of the pin connection holes were marred and damaged; he ordered a new back plane board and another font end pcb.The fse returned to the site once again and replaced the back plane board.Once replaced, he was able to connect the front end board without any issues.Performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.The full event site name is (b)(6).
 
Event Description
It was reported that during during set up of unit.It was discovered that the cardiosave intra-aortic balloon pump (iabp) arterial pressure port was not working.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Additional contact information: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key11865567
MDR Text Key252264948
Report Number2249723-2021-01080
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-